Operating the CMC 356
17
3.3 Regular Test Leads for Safety Sockets
Use the regular test leads of 2.0 m/6 ft. length to connect the
CMC 356
output to other safety sockets of, for example, amplifiers, test objects or to
banana adapters in control cabinets.
3.3.1 Terminal
adapters
The optional CMC Wiring Accessory Package includes flexible terminal
adapters to connect the regular test leads to screw-clamp terminals.
The terminal adapters have blank ends. Therefore, turn off the voltage
before connecting these adapters. Always insert an adapter with its blank
end into the terminal strip first, and fasten it before connecting it to a test
lead.
Regular test lead
or to safety socket,
e.g., at test object.
CMC 356
test set
or amplifier
to terminal strip
↔
Regular test lead
Terminal adapter