Communication Protocol (TRCN-443):
1) Proprietary:
Order:
ESC ID P CR LF
0x1B 0x50 0x0D 0x0A Hexadecimal
The ID is configured at the instrument from 1 to 32.
Answer:
L >
> V V V V V V
N T U
C C C C C C
m A
Example of answer for an un-stable value (for low turbidity 0 – 10NTU)
L > 0.001
N T U
1 2 . 0 0
m A
Example of answer for a stable value (for low turbidity 0 – 10NTU)
L > > 0.001
N T U
1 2 . 0 0
m A
Note: a) When the answer comes a C instead of a L, it means that the equipment is under
calibration function (is being operated in location by the user, executing the calibration
operation at the instrument).
b) When the answer comes an S instead of a L, it means that the equipment is under Set Up
function and it is being operated in location by the user.
2) RS485
– It is a “physical location”, where the proprietary protocol will be “transported”.
As factory default, this communication comes configured as:
Speed = 9600
Parity = none
Number of Bits = 8
Stop Bit = 1
ID = 1
23
13.
Proprietary Communication Protocol