Before initiating the Calibrtion Procedure, preapre the Standard Solutions and have the deionized water
ready. Also clean the Calibration Reservoir using deionized waterat every change of Standard Solution.
TURBIDIMETER
/SET UP
READ
TURBIDIMETER
READ/
CALIBRATE
CAL. 0.3000 NTU?
/
YES
NO
CONFIRM ?
/
YES
NO
PLACE SENSOR
@ 0.3000 NTU
READY ?
================
WAIT
NON CONFORMANCE
STANDARD <ENTER>
CONFIRM STAND.?
/
YES
NO
CAL.:1.0000 NTU?
/
YES
NO
CONFIRM ?
/
YES
NO
PLACE SENSOR
@ 1.0000 NTU
READY ?
================
WAIT
GO TO SAMPLE!
READY ?
================
->8.4390 NTU
S
ta
n
d
a
rd
i
n
c
o
n
fo
rm
a
n
c
e
12.
Equipment Operation - Calibration
21
Press
until
flashes then press
key.
<SEL>
Read
<ENT>
Press
until
flashes then press
key.
<SEL>
Calibrate
<ENT>
Follow intructions as per page 13, dip the Standard
Solution on the value shown, this Std Solution is chosen
during Set Up Mode. Press
until desired opiton
flashes then press
key. User will still have the
option to accept this calibration point or not!
Press
key to confirm.
Place the sensor unit, as per instruciton on page 13, into
the Standard value shown. This Std Solution is chosen
during Set Up Mode. Press
until desired opiton
flashes then press
key.
Press
key to confirm.
The equipment will analyze the Standard Solution and
If it does not match the instrument linarity, it will inform
the user about its quality, but if the user decides to use
this solution, he will be able to do so!
Follow intructions as per page 13, dip the Standard
Solution on the value shown, this Std Solution is chosen
during Set Up Mode. Press
until desired opiton
flashes then press
key. User will still have the
option to accept this calibration point or not!
Press
key to confirm.
Place the sensor unit, as per instruciton on page 13, into
the Standard value sown. Tthis Std Solution is chosen
during Set Up Mode. Press
until desired opiton
flashes then press
key.
Press
key to confirm.
Remove the sensor from the bubble trapper, remove the
Calibration cup. Re-insert the sensor into the trapper,
open the process and proceed with Readings.
<SEL>
<SEL>
<SEL>
<SEL>
<ENT>
<ENT>
<ENT>
<ENT>
<ENT>
<ENT>
<ENT>
<ENT>
S1:» S2:»
-> 8.4390 NTU
Indicates when the
Bargraph was not chosen
during Set Up Mode.