APPLICATION
Analyzer / Transmiter / Turbidity Controller
ANALYZER
Display
Alphanumeric 2 lines x 16 characters
Range
0.0001 to 10 NTU
Resolution
0.0001 (0.0001 thru 5NTU) / 0.001 (5.001 thru 10NTU)
Relative Precision
0.01% (full scale)
Analytical Method
Nephelometric
Wavelength
660 nm
GENERAL
Enclosure
IP-68
Body Material
Cast Aluminum SAE 323(BASE)
ABS (Frontal Lid)
Electrical Connection
Terminal Barr
Cable Inlet
Cale Knockout 3/8” BSP (4)
Installation
Flat Surface
Relative Humidity
20 to 80 %
Electrical Power
90 to 240 VAC (50/60 Hz)
Power Consumption
3.5 VA
Sample Flow
250 to 750 ml/min
Sample Pressure
(Min./max.)
10 / 20 psi
Sample Temperature
5 to 40°C
Weight
5 Kg (equipment and bubble trapper)
TRANSMITER
Analogic Output
2x 4 to 20 mA
Digital Output
RS-485
CONTROLER
Type ON-OFF / PWM
Two,NO (1A / 250 VAC)
Set-Point
from 0 to 100% of scale
Instruction Manual
Installation Hardware
2 aluminum base SAE 323, 2 “U” clamps in SS 316,
Nuts and washers in SS 316
2.
Specifications
ACCESSORIES
3
Summary of Contents for TRCN443