CURRENT CONFIG.?
/
YES
NO
OUTPUT CONFIG.:
mA
MA OUTPUT:
/
ON
OFF
VALUE FOR 4 mA
0.0000NTU
<
>
VALUE FOR 20 mA
5.0000NTU
<
>
BURN-OUT CONFIG:
STAND-BY:
4mA
CALIBRATE mA?
/
YES
NO
CONFIRM?
/
YES
NO
PLACE AMP METER
@ OUTPUT 4-20 mA
READY?
ADJUST 4 mA
<SEL>- <ESC>+
12.
Equipment Operation - Set Up (cont.)
19
User has the option to configure Current.
Press
key to choose the desired option then
press
key to confirm.
<
SELECT>
<ENTER>
User has the option to choose using Output mA.
Press
key to choose the desired option then
press
key to confirm.
<
SELECT>
<ENTER>
User can adjust the value the values for 4 and 20mA
Current Output.
Refer to Page 16 (Note A) for instruction on how to adjust
this value. Resolution will depend upon the Resolution
chosen while at Set Up.
For the Burn Out configuration, user will have three options
as: 4mA, 20mA or Hold.
Press
key until the desired option flashes, then
press to
confirm.
<SEL>
<ENT>
User will be able to calibrate mA output.
Press
key to choose the desired option then
press
key to confirm
<
SELECT>
<ENTER>
Connect Amp Meter to output 4-20mA and it will be possible
to adjust them, then press
<ENTER>
when ready.
Adjust the value as needed by pressing
(to
decrease) or
(to increase) keys, so the outputs
can be adjusted.
key will decrease the value and
key
will increase.
<SELECT
>
<SELECT>
<ESCAPE>
<ESCAPE>
Go to Page 20
ADJUST 20 mA
<SEL>- <ESC>+
SET POINT S2:
20.0000 NTU
<
>
ACTING:
/REVERSE
DIRECT
HYSTERESE:
0.0000NTU
<
>
BURN-OUT CONFIG:
STAND-BY:
HOLD
User can adjust the value for SP 2.
Refer to Page 16 (Note A) for instruction on how to adjust
this value. Resolution will depend upon the Resolution
chosen while at Set Up.
User can choose the Acting desired.
Press
key to choose the desired option then
press
key to confirm
<
SELECT>
<ENTER>
User can adjust the value for Hystereses.
Refer to Page 16 (note A)for instruction on how to adjust
this value. Resolution will depend upon the Resolution
chosen while at Set Up.
For the Burn Out configuration, user will have three options as:
On - Contact is going to stay On all the time, Off - contact will stay
Off all the time or Hold - contact is going to follow the last situation,
before going to Hold status.
Press
key until the desired option flashes, then press
to confirm.
<SEL>
<ENT>
From Page 18
A