4.7.5.8 List of Available Communication Commands:
Table 4.2 List of Communication Commands
Command
Command
Function
List
Response
a=xxx
Device Measurement Mode/Type
b=xxx
CH A & CH B Combined input reading (A-B F or A-B T Mode)
c=xxx
CH A input reading value
d=xxx
CH B input reading value
e=xxx
Your own-defined Name of Experiment/Measurement Task
f=xxx
Unit of Input Value (definable)
*SRP
g=ABC
= 3 bits:
A=1st bit: the transmit active bit
B=2nd bit: the SP1 active bit
C=3rd bit: the SP2 active bit
h=xxx
SP1 (setpoint 1) value (A Batch value if Batch mode used)
i=xxx
SP2 (setpoint 2) value (B Batch value if Batch mode used)
j=xxx
Batch A Number (Number of A Batches)
k=xxx
Batch B Number (Number of B Batches)
l=xxx
A-B mode sign.
M=xxx
User-definable time/date stamp of Input Reading.
*SRR
A-B mode reading
*SRA
Channel A input reading
*SRB
Channel B input reading
*SRX
Reset_A
Reset Channel A counter to 0
*SRY
Reset_B
Reset Channel B counter to 0