Configuring The Meter
34
4.8 SETTING THE SETPOINT 1 DEADBAND
S1.DB
Refer to
Table 6-1
for a summary list of menu configuration.
Setpoint 1 Deadband
S1.DB
is not active unless your meter has dual relay output
capabilities. The LED's will display whether the
S1.DB
is active or not. The Setpoint 1
Default deadband is 0003. To change the deadband (hysteresis) of Setpoint 1, follow
these steps:
1.
Press the
MENU
button until the meter shows
S1.DB
.
2.
Press the
∂
TARE
button. The meter shows the last previously stored 4-digit
number (0000 through 9999) with flashing 4th digit.
3.
Press the
ß
NT/GRS
button to change the value of the flashing digit. If you
continue to press the
ß
NT/GRS
button, the flashing digit's value continues to
change.
4.
Press the
∂
TARE
button to scroll to the next digit.
5.
Press the
MENU
button to store your selection. The meter momentarily shows
STRD
, followed by
S2.DB
(Setpoint 2 Deadband).
Note
☞