4
Configuring The Meter
38
38
4.10.2 Selecting Analog Output as Current or Voltage
1.
Press the
MENU
button until it shows
OT.CF
, then press the
∂
TARE
button twice.
or
Press the
∂
TARE
button from
0.1
.
The meter shows one of the following:
•
O.2=C
(Analog output = current)
(Default)
•
O.2=V
(Analog output = voltage)
2.
Press the
ß
NT/GRS
button to toggle between available choices.
3.
Press the
∂
TARE
button to select analog output or proportional control or press the
MENU
button to store your selection and enter
OT.S.O
(Output Scale and Offset).
4.10.3 Selecting Analog Output or Proportional Control
1.
Press the
MENU
button until it shows
OT.CF
, then press the
∂
TARE
button twice.
or
Press the
∂
TARE
button from
0.2
.
The meter shows one of the following:
•
O.3=N
(Net Reading)
(Default)
•
O.3=G
(Gross Reading)
2.
Press the
ß
NT/GRS
button to toggle between available choices.
3.
Press the
MENU
button to store your selection. The meter momentarily shows
STRD
, followed by
OT.S.O
(Output Scale and Offset).