DMTA-10026-01EN, Rev. D, January 2015
Basic and Multipoint Calibration
77
In order to make accurate measurements with the Magna-Mike 8600, the user must
perform a four-point basic or multipoint (additional points) calibration for the target
ball, disk, or wire that is being used. The points used in the calibration will be a
“BALL OFF” point, a “BALL ON” point (thickness equal to zero), a “THIN SHIM”
point, and a “THICK SHIM” point. The values for the “THIN SHIM” and “THICK
SHIM” calibration points will vary depending on the wear tip used, target ball, disk,
or wire selected. The user may also choose to add up to eight additional calibration
points in order to fine-tune the calibration curve and improve accuracy in the
thickness range where measurements will be made.
4.7.1
Selecting a Target and Wear Tip
When using the standard target balls and the standard wear tip, the Magna-Mike 8600
can normally automatically recognize the proper ball during the calibration for the
86PR-1 and 86PR-2 probes. The Magna-Mike 8600 will display the recognized target
on the calibration screen. When using the 86PR-3 probe (and sometimes with the
other two probes as well), the instrument cannot recognize the correct target ball size
and the user will have to manually choose the target ball in the TARGET SELECT
field. When using the target disk, the target wire, or the chisel wear tip, the user must
also manually select the proper target and wear tip in the MEASUREMENT SETUP
screen.
To manually select the proper target and wear tip
1.
In the measurement screen, press
[SET UP]
, and then highlight
MEASUREMENT.
2.
In the
MEASUREMENT
submenu, select
TARGET SELECT
, and then press
[ENTER]
.
3.
In the
MEASUREMENT SETUP
screen (see Figure 4-5 on page 78), proceed as
follows:
When using the 86PR-1 or 86PR-2 and one of the target balls, select
Auto Ball
.
OR
When using the 86PR-3 or one of the target disks or wires, select
1/16 BALL
,
1/8 BALL
,
0.5 DISK
,
V DISK
,
0.045 WIRE
, or
0.026 WIRE
.
Note that the targets available to select are probe dependent.
4.
Highlight
WEARTIP SELECT
, and then choose
STANDARD
or
CHISEL
.
Note that only the
STANDARD
wear tip setting is available for the 86PR-3 probe.
5.
Press
[MEAS]
to return to the measurement screen.
Summary of Contents for Magna-Mike 8600
Page 7: ...DMTA 10026 01EN Rev D January 2015 Table of Contents vii List of Tables 169 Index 171...
Page 8: ...DMTA 10026 01EN Rev D January 2015 Table of Contents viii...
Page 10: ...DMTA 10026 01EN Rev D January 2015 List of Abbreviations x...
Page 28: ...DMTA 10026 01EN Rev D January 2015 Introduction 18...
Page 70: ...DMTA 10026 01EN Rev D January 2015 Chapter 3 60...
Page 112: ...DMTA 10026 01EN Rev D January 2015 Chapter 5 102...
Page 120: ...DMTA 10026 01EN Rev D January 2015 Chapter 6 110...
Page 146: ...DMTA 10026 01EN Rev D January 2015 Chapter 7 136...
Page 160: ...DMTA 10026 01EN Rev D January 2015 Chapter 8 150...
Page 164: ...DMTA 10026 01EN Rev D January 2015 Chapter 9 154...
Page 168: ...DMTA 10026 01EN Rev D January 2015 Appendix A 158...
Page 178: ...DMTA 10026 01EN Rev D January 2015 List of Figures 168...
Page 180: ...DMTA 10026 01EN Rev D January 2015 List of Tables 170 Table 31 Calibration kits 163...
Page 186: ...DMTA 10026 01EN Rev D January 2015 Index 176 WINXL about 137...