DMTA-10026-01EN, Rev. D, January 2015
Basic and Multipoint Calibration
61
4. Basic and Multipoint Calibration
Olympus recommends regular calibration (with regular verification) of the Magna-
Mike 8600 to ensure consistency and accuracy of thickness readings. This chapter
outlines how to choose probes, target balls, target disks, wire targets, when to
calibrate, and how to calibrate.
4.1
Probe Types
The Magna-Mike 8600 can be used with three different probes:
•
Standard straight probe (Olympus P/N: 86PR-1 [U8470020])
•
Right angle probe (Olympus P/N: 86PR-2 [U8470028])
•
Low-profile articulating probe (Olympus P/N: 86PR-3 [Q7800004])
The 86PR-1 and 86PR-2 probes have the same measurement and maximum thickness
capabilities, but the 86PR-2 is shorter, has a handle, and is easier to use when the
probe is being used outside of the probe stand. The 86PR-2 probe (see Figure 4-2 on
page 62) is useful for special application where the longer 86PR-1 probe (see
Figure 4-1 on page 62) may be too large to fit into the measurement location.
The third probe (86PR-3) is a special low-profile probe designed to be as small as
possible for applications with very limited access space at the measurement location
(see Figure 4-3 on page 63). However, this probe has a more limited thickness range.
The head of the probe can be adjusted and locked at different angles. The 86PR-3
probe’s maximum measurable thickness is up to 4.06 mm (0.160 in.), depending on
the target being used.
Summary of Contents for Magna-Mike 8600
Page 7: ...DMTA 10026 01EN Rev D January 2015 Table of Contents vii List of Tables 169 Index 171...
Page 8: ...DMTA 10026 01EN Rev D January 2015 Table of Contents viii...
Page 10: ...DMTA 10026 01EN Rev D January 2015 List of Abbreviations x...
Page 28: ...DMTA 10026 01EN Rev D January 2015 Introduction 18...
Page 70: ...DMTA 10026 01EN Rev D January 2015 Chapter 3 60...
Page 112: ...DMTA 10026 01EN Rev D January 2015 Chapter 5 102...
Page 120: ...DMTA 10026 01EN Rev D January 2015 Chapter 6 110...
Page 146: ...DMTA 10026 01EN Rev D January 2015 Chapter 7 136...
Page 160: ...DMTA 10026 01EN Rev D January 2015 Chapter 8 150...
Page 164: ...DMTA 10026 01EN Rev D January 2015 Chapter 9 154...
Page 168: ...DMTA 10026 01EN Rev D January 2015 Appendix A 158...
Page 178: ...DMTA 10026 01EN Rev D January 2015 List of Figures 168...
Page 180: ...DMTA 10026 01EN Rev D January 2015 List of Tables 170 Table 31 Calibration kits 163...
Page 186: ...DMTA 10026 01EN Rev D January 2015 Index 176 WINXL about 137...