![Olympus Magna-Mike 8600 User Manual Download Page 68](http://html1.mh-extra.com/html/olympus/magna-mike-8600/magna-mike-8600_user-manual_3943437068.webp)
DMTA-10026-01EN, Rev. D, January 2015
Chapter 3
58
3.6
Changing the Thickness Resolution
The resolution of the Magna-Mike 8600 is set by default to AUTO, where the number
of decimal places shown on the display varies, depending on the thickness being
measured.
While in automatic resolution, the measurements between 0.000 mm (0.000 in.) and
4.06 mm (0.160 in.) are displayed at high resolution (0.001 mm or 0.0001 in.).
Measurements above 4.06 mm (0.160 in.) are displayed in standard resolution
(0.01 mm or 0.001 in.).
For the 1.59 mm (1/16 in.) target ball, the automatic resolution changes from HIGH to
STANDARD at 2.03 mm (0.080 in.).
You can change the thickness measurement resolution, meaning the number of digits
shown to the right of the decimal point. The resolution selection affects all displays
and data output of values with thickness units. This includes the measured thickness,
the differential reference value, and the alarm set points. The highest thickness
resolution is 0.001 mm (0.0001 in.).
You can reduce the resolution in some applications where the extra precision of the
last digit is not required, or where rough outside or inside surfaces make the last
thickness display digit unreliable.
To change the thickness measurement resolution
1.
In the measurement screen, press
[SET UP]
, and then highlight
MEASUREMENT
.
2.
In the
MEASUREMENT
submenu, select
RESOLUTION
.
3.
In the
MEASUREMENT SETUP
screen (see Figure 3-2 on page 53), set the
RESOLUTION
to the desired option:
•
AUTO
(default): Automatically changes between
STANDARD
and
HIGH
resolution based on the thickness being measured. Thicknesses below
4.06 mm (0.160 in.) are displayed at
HIGH
resolution, while thicknesses
above 4.06 mm (0.160 in.) are displayed in
STANDARD
resolution
•
STANDARD
: 0.01 mm or 0.001 in.
NOTE
Summary of Contents for Magna-Mike 8600
Page 7: ...DMTA 10026 01EN Rev D January 2015 Table of Contents vii List of Tables 169 Index 171...
Page 8: ...DMTA 10026 01EN Rev D January 2015 Table of Contents viii...
Page 10: ...DMTA 10026 01EN Rev D January 2015 List of Abbreviations x...
Page 28: ...DMTA 10026 01EN Rev D January 2015 Introduction 18...
Page 70: ...DMTA 10026 01EN Rev D January 2015 Chapter 3 60...
Page 112: ...DMTA 10026 01EN Rev D January 2015 Chapter 5 102...
Page 120: ...DMTA 10026 01EN Rev D January 2015 Chapter 6 110...
Page 146: ...DMTA 10026 01EN Rev D January 2015 Chapter 7 136...
Page 160: ...DMTA 10026 01EN Rev D January 2015 Chapter 8 150...
Page 164: ...DMTA 10026 01EN Rev D January 2015 Chapter 9 154...
Page 168: ...DMTA 10026 01EN Rev D January 2015 Appendix A 158...
Page 178: ...DMTA 10026 01EN Rev D January 2015 List of Figures 168...
Page 180: ...DMTA 10026 01EN Rev D January 2015 List of Tables 170 Table 31 Calibration kits 163...
Page 186: ...DMTA 10026 01EN Rev D January 2015 Index 176 WINXL about 137...