![Olympus Magna-Mike 8600 User Manual Download Page 106](http://html1.mh-extra.com/html/olympus/magna-mike-8600/magna-mike-8600_user-manual_3943437106.webp)
DMTA-10026-01EN, Rev. D, January 2015
Chapter 5
96
Figure 5-5 Example of a HIGH alarm indicator
The datalogger records an alarm condition in the second status box for all stored
measurements. An
A
indicates the alarm mode, an
L
indicates a low alarm condition,
and an
H
indicates a high alarm condition.
To set the alarm
1.
In the measurement screen, press
[SET UP]
, and then highlight
ALARM
.
2.
In the
ALARM
submenu, select
ENABLE
,
LOW ALARM SET POINT
, or
HIGH
ALARM SET POINT
.
3.
In the
ALARM SETUP
screen (see Figure 5-6 on page 97):
a)
Set
ENABLE
to
ON
to activate the alarm function.
b)
Set the
LOW ALARM SET POINT
, and then press
[ENTER]
.
c)
Set the
HIGH ALARM SET POINT
.
d)
Press
[MEAS]
to return to the measurement screen.
High alarm indicator
Thickness reading (turns
red during an alarm
condition)
Summary of Contents for Magna-Mike 8600
Page 7: ...DMTA 10026 01EN Rev D January 2015 Table of Contents vii List of Tables 169 Index 171...
Page 8: ...DMTA 10026 01EN Rev D January 2015 Table of Contents viii...
Page 10: ...DMTA 10026 01EN Rev D January 2015 List of Abbreviations x...
Page 28: ...DMTA 10026 01EN Rev D January 2015 Introduction 18...
Page 70: ...DMTA 10026 01EN Rev D January 2015 Chapter 3 60...
Page 112: ...DMTA 10026 01EN Rev D January 2015 Chapter 5 102...
Page 120: ...DMTA 10026 01EN Rev D January 2015 Chapter 6 110...
Page 146: ...DMTA 10026 01EN Rev D January 2015 Chapter 7 136...
Page 160: ...DMTA 10026 01EN Rev D January 2015 Chapter 8 150...
Page 164: ...DMTA 10026 01EN Rev D January 2015 Chapter 9 154...
Page 168: ...DMTA 10026 01EN Rev D January 2015 Appendix A 158...
Page 178: ...DMTA 10026 01EN Rev D January 2015 List of Figures 168...
Page 180: ...DMTA 10026 01EN Rev D January 2015 List of Tables 170 Table 31 Calibration kits 163...
Page 186: ...DMTA 10026 01EN Rev D January 2015 Index 176 WINXL about 137...