![Olympus Magna-Mike 8600 User Manual Download Page 53](http://html1.mh-extra.com/html/olympus/magna-mike-8600/magna-mike-8600_user-manual_3943437053.webp)
DMTA-10026-01EN, Rev. D, January 2015
Instrument Overview
43
Figure 1-22 Instrument stand
1.6.3.3
O-Ring Gasket and Membrane Seals
The Magna-Mike 8600 contains seals that are used to protect the instrument internal
hardware from the environment. These include:
•
Battery compartment cover seal
•
I/O door seal
•
Membrane vent
These seals must be properly maintained to assure environmental durability.
Instrument seals are evaluated and replaced as needed during the instrument annual
calibration. This should be performed at an authorized Olympus service center.
1.6.3.4
Display Protection
The Magna-Mike 8600 includes a clear-plastic sheet protecting the instrument display
window. Olympus strongly recommend to leave this protection sheet in place.
Replacements are available in packages of ten sheets (Olympus P/N: 600-DP
[U8780297]).
Summary of Contents for Magna-Mike 8600
Page 7: ...DMTA 10026 01EN Rev D January 2015 Table of Contents vii List of Tables 169 Index 171...
Page 8: ...DMTA 10026 01EN Rev D January 2015 Table of Contents viii...
Page 10: ...DMTA 10026 01EN Rev D January 2015 List of Abbreviations x...
Page 28: ...DMTA 10026 01EN Rev D January 2015 Introduction 18...
Page 70: ...DMTA 10026 01EN Rev D January 2015 Chapter 3 60...
Page 112: ...DMTA 10026 01EN Rev D January 2015 Chapter 5 102...
Page 120: ...DMTA 10026 01EN Rev D January 2015 Chapter 6 110...
Page 146: ...DMTA 10026 01EN Rev D January 2015 Chapter 7 136...
Page 160: ...DMTA 10026 01EN Rev D January 2015 Chapter 8 150...
Page 164: ...DMTA 10026 01EN Rev D January 2015 Chapter 9 154...
Page 168: ...DMTA 10026 01EN Rev D January 2015 Appendix A 158...
Page 178: ...DMTA 10026 01EN Rev D January 2015 List of Figures 168...
Page 180: ...DMTA 10026 01EN Rev D January 2015 List of Tables 170 Table 31 Calibration kits 163...
Page 186: ...DMTA 10026 01EN Rev D January 2015 Index 176 WINXL about 137...