87
IPLEX NX
This mode allows for measuring the distance from a measurement point to the reference plane
specified by three reference points. It indicates depth and height.
Height is indicated by a positive value and depth is indicated by a negative value.
This mode allows for measuring the total length of multiple lines, each of which is specified by
two measurement points. When the specification of the last measurement point causes the
first-drawn line and last-drawn line to intersect, the area of the figure closed by the lines will
be measured. However, note that the measured area does not represent the precise area of
the measurement target but is only an approximate value obtained by dividing the figure into
triangles. Up to 20 measurement points can be specified per measurement session.
<Depth>
<Area/Lines>
<Depth>
Depth or
height
Measurement
point
Reference
point
Reference
point
Reference
point
Reference
plane
<Area/Lines>
1
6
2
3
4
5
Measurement
point
Measurement
point
Measurement point
Measurement
point