135
Chapter 8 Storage and maintenance
WARNING
•
Before storing the main unit in the case, be sure to remove the battery and turn the main unit
off. Otherwise, the equipment may overheat and cause a fire.
CAUTION
•
Stow the control unit, insertion tube, and main unit in the correct order. If they are not stored
in the correct order, the insertion tube may be damaged.
•
Let the distal end of the insertion tube cool before stowing it in the case.
•
Do not apply excessive force when stowing the insertion tube into the cushioned slot.
Otherwise, the insertion tube may be damaged.
•
Make sure to remove twist from the insertion tube when storing it.
•
Make sure that the case’s latch is locked properly. If it is locked improperly, the top cover may
open during transport.
•
When closing the top cover, be careful that your hand, the strap, insertion tube or cables are
not pinched.
•
When returning the main unit to its case, open the case in a stable place before repacking the
equipment.
•
Use only the designated case. Failure to do so may cause damage or malfunction of the
insertion tube and/or main unit.
NOTE
•
Refer to the storage process label on the inside of the top cover of the carrying case for the
storage procedure.
Locking the case
The top cover of the case may open if the case is hit while being moved. Use the
provided key to lock the case during transport.
CAUTION
•
Lock the case at both of the two locations.
Summary of Contents for IPLEX LX Series
Page 2: ......
Page 8: ...vi For 8 5 mm type insertion tube 157 Appendix 158 System chart 158...
Page 10: ...2 3 Warning plate...