123
Chapter 6 Measurement procedures (LX only)
Exiting from measurement
Press the [LIVE/GAIN] button to display the exit measurement screen and then
select “OK” to exit measurement. Select “CANCEL” to return to the stereo
measurement screen.
Select “EXIT” in the stereo measurement menu to exit measuring and return to
the screen before measurement is activated.
In case accuracy cannot be achieved
(1) Bring the insertion tube as close as possible to the object so that the
screen is filled with the object.
(2) Measure the length so that it is parallel with the screen whenever
possible. While the resolution of measuring is high when the
measuring direction is parallel with the screen, it gets low when
measuring direction becomes closer to depthwise. Therefore, the
measurement can be made more accurate by measuring a parallel
object to the screen.
(3) Change the point of view and capture an image again if the objects
have similar patterns around measurement points have few surface
characteristics or if the image is blurry or reflective.
(4) Check that the optical adapter is attached properly to the insertion
tube.
(5) Check that there is no dirt between the optical adapter and insertion
tube, and attach the optical adapter again.
(6) If the accuracy still cannot be improved, execute “NEW STEREO
ADAPTER” again (see page 108).
Exit measurement confirmation window