45
GX53
6
6
Replacement of optical parts
1
Removing the eyepiece
2
Attaching the eyepiece micrometer
Pull out the eyepiece a from the eyepiece sleeve b .
1
a
b
You can use the eyepiece micrometer to check the size of the observed
image or the center position of the observation field of view, etc. For
the eyepiece micrometer types attachable to your eyepiece, contact
Olympus.
Attachable eyepiece micrometer
Size
Ø24 mm Thickness 1.5 mm
Hold the eyepiece frame, and rotate the built-in reticle holder a in
the arrow direction to remove it.
Depending on the case, the reticle holder may be tightened
too firmly and it cannot be rotated. In this case, if you grab
the reticle holder too firmly, it will be deformed and cannot be
rotated furthermore. Grab around the reticle holder lightly with
equal force to rotate it or push it to the rubber sheet placed
on the desk to rotate and remove it.
Insert the eyepiece micrometer b in the removed reticle holder a
with the display surface of the eyepiece micrometer facing down.
Screw in the reticle holder a to the bottom of the eyepiece.
·
Be careful not to touch the lens with your fingers during
working.
·
Do not tighten the reticle holder with excess force.
1
TIP
2
3
NOTE
1
2
6-1 Replacing the eyepiece
b
a
1