Attaching the illumination column
Loosen the clamping screws (4 positions) using the Allen wrench
provided with the microscope frame to remove the illumination column
dummy block attached to the microscope frame.
Align the guide holes on the illumination column side with the
positioning pins a (2 positions) protruded from the microscope frame,
and fit the illumination column from above.
Holding the illumination column with your hand, secure the illumination
column to the microscope frame with the provided hex socket screws
b (4 positions) using the provided Allen wrench.
1
2
3
For attaching procedures, see “Replacing the halogen bulb" on page 48.
Loosen the lamp housing clamping screws a (2 positions) of the
illumination column using the Allen screwdriver and insert the lamp
housing b .
Make sure that the lamp housing is attached correctly without tilting
as shown in the picture and tighten the clamping screws a .
1
2
a
a
b
This picture shows the procedure to attach the
halogen lamp housing (U-LH100L-3). The same
procedure applies when attaching the LED light
source for reflected light observation (BX3M-LEDT).
b
12
Attaching the halogen bulb for transmitted
light observation
13
Attaching the lamp housing for transmitted
light observation