59
GX53
Meaning of notations in the following tables
Indicates the lighting state of the indicator of the control box (BX3M-CBFM).
: Turns ON.
: Blinks. :Turns OFF. : Varies depending on state.
Problem
Cause
Remedy
Page
7. Control box (BX3M-CBFM) and hand switch (BX3M-HS)
a) Even though the objective is changed by rotating the revolving nosepiece, the OB indicator of the hand switch does not
turn ON.
The power of the control box
(BX3M-CBFM) is OFF.
Set the main switch to (OFF) and
connect the control box to the AC
adapter of the control box again.
Then, set the main switch to (ON) .
38, 76
The hand switch (BX3M-HS) is not
connected.
Set the main switch to (OFF) and
connect the hand switch (BX3M-HS).
38, 76
The hand switch (BX3M-HS) is
damaged.
Contact Olympus.
—
The cable of the revolving
nosepiece or GX-IFRES is not
connected.
Set the main switch to (OFF) and
connect the cable of the revolving
nosepiece or GX-IFRES.
38, 68, 74
The objective is not correctly
engaged in the light path.
Set it to the click position properly.
20
For troubleshooting of the control box (BX3M-CBFM) not described in the above table, refer to the instruction manual for the
control box (BX3M-CBFM)
Repair request
If you cannot solve the problems even though taking actions described in Troubleshooting, please contact Olympus
for assistance. At that time, please tell them the following information as well.
· Product name and abbreviation (Example: Stage GX-SVR)
· Product number
· Problem