20
4-2 Adjusting the tension of the coarse focusing knob
The tension of the coarse focusing knob can be adjusted. Insert
the tip of a large flat-blade screwdriver into the groove b on the
tension adjustment ring a and rotate the ring. Rotating the coarse
focusing knob in the arrow direction shown in the picture increases
the tension and rotating it in the opposite direction decreases the
tension respectively.
Never apply the excessive force for rotating the tension
adjustment ring
a
in the arrow direction shown in the
picture. The ring may be damaged..
If the revolving nosepiece descends by its own weight or the focus
obtained with the fine focusing knob c is lost soon, the tension is
set too loose. In this case, turn the tension adjustment ring a in the
arrow direction to increase the tension of the coarse focusing knob.
1
NOTE
2
You can adjust the space between the stage and the umbra shield.
Adjust the space according to the size of the container to be placed on
the stage, if necessary.
Loosen the umbra shield knob a . Then, move the umbra shield along
with the umbra shield column to adjust the position.
Decide the height of the umbra shield and tighten the umbra shield
knob a .
1
2
4-4 Adjusting the umbra shield (ambient light shielding plate) position (Fluorescence set)
4-3 Operating procedures of the transmitted light illumination filter (option)
More effective observation and image acquisition can be made by using filters matching to the purposes.
For procedures to attach the filter to the microscope frame CKX53SF, see page 41.
Filter
Specifications
Intended use
43IF550W45
Monochrome contrast filter (green)
· This filter is used by attaching to the filter
holder for transmitted light illumination.
· The contrast becomes higher.
45-ND25
Filter for light intensity adjustment
(Transmittance: 25 %)
· By attaching this filter to the pre-centered
phase contrast slider CKX3-SLP, the
changes in the brightness can be reduced
when changing from the phase contrast
observation to the bright field observation.
a
c
b
a