31
BX61WI
Fig. 43
Fig. 44
c) Combination using C-mount adapter IR system (visible light to 1000 nm)
C-mount adapter IR system
U-TVCAC
U-PMTVC2XIR
U-PMTVC4XIR
C-mount TV camera
(IR specification model)
d) Combination using U-CA or U-ECA intermediate tube
One of these intermediate tubes can be used only in combination with the U-ETR3 or U-TR30IR trinocular tube. The TV
adapter used in this combination should be one of that used in a), b) or c).
2
Attaching the IR Modules
(Figs. 43 & 44)
Thermal Reflection Filter 45SCF
1. Remove the collector lens of the U-LH10IR lamp housing
1
by loosen-
ing the 3 clamping screws
2
with an Allen wrench (width across flats of
2.5 mm).
2. While positioning the 45SCF filter
3
so that the arrow on its frame points
in the opposite direction of the lamp housing, insert the filter in the lamp
housing, and clamp by tightening the ring spring
4
provided with the
filter.
3. Attach the collector lens to the original position.
IR Filter 23BP775 or 32IR900
}Be sure to insert the 32BP775 or 32IR900 IR filter in the filter slider below
the filter turret. (For the mounting method, see page 15.)
# If the IR filter is inserted above the polarizer in the filter turret, the
polarizer will be burnt.
Other IR Modules
Also replace other required modules with the IR modules (see page 30).
1
2
3
4