DMTA-10009-01EN [U8778346], Rev. D, November 2016
Initial Setup
57
Figure 4
‑
7 The range of the waveform display
4.5.1
Selecting the Range Value
There are fixed ranges available for each transducer frequency. The available ranges
are also dependent on material sound velocity. These selectable ranges let you adjust
the thickness span of the waveform display to only show the thickness range being
measured and thus, obtain maximum waveform resolution for each application. The
range setting affects the waveform display only. You can still make measurements
even when the display range does not show the detected echo from which the
thickness is measured. You cannot manually set the range while the zoom is active.
To select the range value
1.
From the measurement screen, press
[RANGE]
.
The waveform range changes to the next higher available range.
2.
Continue to press
[RANGE]
until you obtain the desired range.
The range value recycles to the minimum range value following the maximum
value.
Selectable range value
Adjustable delay value
Units
Summary of Contents for 38DL PLUS
Page 6: ...DMTA 10009 01EN U8778346 Rev D November 2016 Table of Contents vi...
Page 8: ...DMTA 10009 01EN U8778346 Rev D November 2016 List of Abbreviations viii...
Page 24: ...DMTA 10009 01EN U8778346 Rev D November 2016 Important Information Please Read Before Use 16...
Page 26: ...DMTA 10009 01EN U8778346 Rev D November 2016 Introduction 18...
Page 46: ...DMTA 10009 01EN U8778346 Rev D November 2016 Chapter 2 38...
Page 56: ...DMTA 10009 01EN U8778346 Rev D November 2016 Chapter 3 48...
Page 102: ...DMTA 10009 01EN U8778346 Rev D November 2016 Appendix 94...
Page 106: ...DMTA 10009 01EN U8778346 Rev D November 2016 List of Tables 98...