DMTA-10009-01EN [U8778346], Rev. D, November 2016
Powering the 38DL PLUS
37
6.
At the back of the instrument, unscrew the four screws of the battery
compartment cover (see Figure 2-4 on page 36).
7.
Remove the battery compartment cover.
8.
Remove the battery and carefully disconnect the battery connector.
9.
Connect the new battery in the battery compartment.
10. Ensure that the gasket of the battery compartment cover is clean and in good
condition.
11. Reinstall the battery compartment cover at the back of the instrument, and then
tighten the four screws.
12. Reinstall the protective rubber boot and the hand strap.
13. Press
[ON/OFF]
to turn on the 38DL PLUS.
14. To answer the question appearing at the bottom of the screen (see Figure 2-5 on
Select
Li
‑
ion
when using a 38-BAT battery.
OR
Select
NiMH
or
Alkaline
when using four NiMH AA batteries or four
Alkaline AA batteries with the AA battery holder.
Figure 2
‑
5 Selecting the new battery type
When replacing the rechargeable 38-BAT battery, fully charge the battery to ensure
the accuracy of the estimated remaining battery charge shown by the power indicator
(see “Power Indicator” on page 31).
NOTE
Summary of Contents for 38DL PLUS
Page 6: ...DMTA 10009 01EN U8778346 Rev D November 2016 Table of Contents vi...
Page 8: ...DMTA 10009 01EN U8778346 Rev D November 2016 List of Abbreviations viii...
Page 24: ...DMTA 10009 01EN U8778346 Rev D November 2016 Important Information Please Read Before Use 16...
Page 26: ...DMTA 10009 01EN U8778346 Rev D November 2016 Introduction 18...
Page 46: ...DMTA 10009 01EN U8778346 Rev D November 2016 Chapter 2 38...
Page 56: ...DMTA 10009 01EN U8778346 Rev D November 2016 Chapter 3 48...
Page 102: ...DMTA 10009 01EN U8778346 Rev D November 2016 Appendix 94...
Page 106: ...DMTA 10009 01EN U8778346 Rev D November 2016 List of Tables 98...