ModBus Interface
iTrans Fixed Point Single/Dual Gas Monitor
6-4
iTrans User Manual
(P/N: 77023554-1)
Addr
Inst
R/W
Host
R/W
Range
Description
40105
W
R/W
$0000 to $FFFF
Instrument Mode
Holds code for current mode of instrument.
Possible working modes of instrument are listed
below.
$0001
Normal
$0002
Calibration
$0003
Warm-up
$0006
Zeroing
$0008
Fault
$0009
Reset
Examples:
Sensor in zero fault = $0008
Sensor zeroing
= $0006
40106
W
R
$0000 to $FFFF
Status Bits
Holds 16 bits of status for various parameters in
the instrument. A bit value of “1” indicates that
the associated fault condition is present.
Bit 15 = current loop open
Bit 14 = current loop shorted
Bit 13 = power fault
Bit 12 = 5 volt fault
Bit 11 = missing sensor
Bit 10 =
(not defined)
:
Bit 6
=
(not defined)
Bit 5
= zero fault
Bit 4
= calibration fault
Bit 3
= over-range
Bit 2
= failed sensor
Bit 1
= high alarm
Bit 0
= low alarm
Examples:
Missing sensor = Bit 11 is set
= $0800
Power fault and
failed sensor
= Bits 13 and 2 set = $2004
40115
W
R
Last Alarm Date (mmdd)
Holds the month and day when the instrument
had the last alarm.
Summary of Contents for iTrans
Page 13: ...iTrans Fixed Point Single Dual Gas Monitor Introduction P N 77023554 1 iTrans User Manual 1 7...
Page 14: ......
Page 36: ......
Page 54: ...Operation iTrans Fixed Point Single Dual Gas Monitor 5 18 iTrans User Manual P N 77023554 1...
Page 78: ...Appendix D iTrans Fixed Point Single Dual Gas Monitor D 2 iTrans User Manual P N 77023554 1...
Page 80: ...Appendix E iTrans Fixed Point Single Dual Gas Monitor E 2 iTrans User Manual P N 77023554 1...
Page 86: ......
Page 87: ......