iTrans Fixed Point Single/Dual Gas Monitor
Operation
(P/N: 77023554-1)
iTrans User Manual
5-11
The 4-20 mA setpoint is designated
with a “4” displayed in status bit and
the current high end range next to it.
To change the range, press the “
”
button
during
the
10-second
countdown. If you do not press “
”
during the 10-second countdown, the
will return to the Normal
Operating Mode. If you initiate the
4-20 mA range option, the status bit
will start to flash and the
range setpoint can be changed by
using the “
” and “
” keys.
When the desired value is reached,
press the “
” key. If the value is not
saved
before
the
time-out,
the
will go back to the
Programming Mode.
Figure 5-18. Changing the Analog
Output Upper Value
5.6.6. Set System Time – Minute
The system’s clock minute setting is
designated with a “1” in the status
bit and current value next to it. To
change the minutes, press the “
”
button
during
the
10-second
countdown. If you do not press “
”
during the 10-second countdown, the
will return to the Normal
Operating Mode. If you initiate the
minutes option, the status bit will
start to flash and the
minute can be changed by using the
“
” and “
” keys.
Figure 5-19. Setting System Time
(Minutes)
Summary of Contents for iTrans
Page 13: ...iTrans Fixed Point Single Dual Gas Monitor Introduction P N 77023554 1 iTrans User Manual 1 7...
Page 14: ......
Page 36: ......
Page 54: ...Operation iTrans Fixed Point Single Dual Gas Monitor 5 18 iTrans User Manual P N 77023554 1...
Page 78: ...Appendix D iTrans Fixed Point Single Dual Gas Monitor D 2 iTrans User Manual P N 77023554 1...
Page 80: ...Appendix E iTrans Fixed Point Single Dual Gas Monitor E 2 iTrans User Manual P N 77023554 1...
Page 86: ......
Page 87: ......