System Wiring
iTrans Fixed Point Single/Dual Gas Monitor
4-4
iTrans User Manual
(P/N: 77023554-1)
NOTE:
When not using 4-20 mA outputs, use the supplied resistors to connect
CH-1 and CH-2 to GND. If these resistors are not connected and the 4-20 mA
outputs are not used, a “P” will appear on the display, indicating an open loop
condition.
4.5. Sensor Wiring (J3)
Connect the
sensor wires (for on-board, remote or stand-alone) to the
appropriate wiring terminals as follows.
24 V:
Red wire from sensor head
(Red wire on BBIR)
485A:
Yellow wire from sensor head
(White wire on BBIR)
485B:
Black wire from sensor head
(Green wire on BBIR)
GND:
Green wire from sensor head
(Black wire on BBIR)
NOTE:
Shielding from either the controller or remote sensors should be bonded
to the enclosure screw located inside the
.
NOTE:
The 24 V terminal supplies 24 VDC to the sensor for power. This
terminal should not be connected to the output of a 24 VDC power supply.
Figure 4-3. Sensor Connector J3 on the
NOTE:
For dual-sensor configurations, place both of the same colored wires in
the appropriate terminal block and firmly tighten.
Summary of Contents for iTrans
Page 13: ...iTrans Fixed Point Single Dual Gas Monitor Introduction P N 77023554 1 iTrans User Manual 1 7...
Page 14: ......
Page 36: ......
Page 54: ...Operation iTrans Fixed Point Single Dual Gas Monitor 5 18 iTrans User Manual P N 77023554 1...
Page 78: ...Appendix D iTrans Fixed Point Single Dual Gas Monitor D 2 iTrans User Manual P N 77023554 1...
Page 80: ...Appendix E iTrans Fixed Point Single Dual Gas Monitor E 2 iTrans User Manual P N 77023554 1...
Page 86: ......
Page 87: ......