Troubleshooting
iTrans Fixed Point Single/Dual Gas Monitor
8-2
iTrans User Manual
(P/N: 77023554-1)
Symptom
Problem
Solution
Reading drifts by 10
counts over a short
time period (in a
stable temperature
environment)
Sensor has failed
Electronics module has
failed
Replace sensor and
calibrate.
Replace electronics module
and calibrate.
In calibration, LED
displays wrong value.
Sensor has failed
Electronics module has
failed
Replace sensor and
calibrate.
Replace electronics module
and calibrate.
Reed Switch does not
work
Electronics module has
failed
Reed Switch is damaged
Replace electronics module
and calibrate.
Replace the reed switch.
“P” appears on the
display
Open loop on a 4-20 mA
channel
Place a 100-Ohm load
resistor from the mA output
pin to ground.
“U” appears on the
display
4-20 mA signal is in over
range
Ensure the sensor is
working properly and the
4-20 mA is scaled correctly.
8.3. Fault Codes
Table 8-2. Fault Codes
Fault
Code
LED
Display
Status
Bit
4-20 mA
Output
Description
0FFF
0.
Flashing
1 mA
Zeroing error – Recover after
calibrating
CFFF
C.
Flashing
1 mA
Calibration error – Recover after
calibrating or replacing the sensor
1FFF
1.
Flashing
1 mA
SMART sensor error
2FFF
2.
Flashing
1 mA
Sensor error
Summary of Contents for iTrans
Page 13: ...iTrans Fixed Point Single Dual Gas Monitor Introduction P N 77023554 1 iTrans User Manual 1 7...
Page 14: ......
Page 36: ......
Page 54: ...Operation iTrans Fixed Point Single Dual Gas Monitor 5 18 iTrans User Manual P N 77023554 1...
Page 78: ...Appendix D iTrans Fixed Point Single Dual Gas Monitor D 2 iTrans User Manual P N 77023554 1...
Page 80: ...Appendix E iTrans Fixed Point Single Dual Gas Monitor E 2 iTrans User Manual P N 77023554 1...
Page 86: ......
Page 87: ......