19
9.2
Alarm Icons and meanings
LEO can display different alarms, which are identified by the Alarm Icons displayed on the Homepage
during an alarm condition.
The
Yellow Warning Alarm Icon
(15 in Figure 19) indicates that the current CO2 or O2
value is out of the range defined in the alarm subpage (see paragraph 9.3.4) during the
Transient Time. The meter
doesn’t
trigger in Alarm and no action is needed. As soon as the CO2
or O2 value is within the defined range, the
Yellow Warning Alarm Icon
disappears. If after the
Transient Time the CO2 or O2 value is still out of the range defined by the operator, the
Yellow
Warning Alarm Icon
changes to
Red Warning Alarm Icon
.
Tip
►
If the temperature sensors are connected to LEO, the Yellow Warning Alarm Icon
appears next to T1 and/or T2 during an alarm condition.
The
Red Warning Alarm Icon
indicates that the current CO2 or O2 value is out of the
range defined in the alarm subpage (see paragraph 9.3.4) after the Transient Time. The meter
triggers in Alarm.
Tip
►
If the temperature sensors are connected to LEO, the Red Warning Alarm Icon
appears next to T1 and/or T2 during an alarm condition.
The
Humidity Alarm Icon
indicates that the current humidity value in LEO is higher than
80% (for example the Moisture Trap is full). The meter triggers an Alarm. The Alarm Icon
disappears when the humidity returns to a value lower than 80%. This alarm is always active
during all type of measurements. If the alarm occurs during a measurement in Aspiration Mode,
stop the Pump. Failing to do so will expose the CO2 O2 sensor to moisture damaging it.
9.3
Settings
Press on
Settings
icon
(see Figure 20 a) to enter the
Settings
menu, as shown in Figure 20 b. The
settings menu has eight items:
1.
Device. See paragraph 9.3.1
2.
Calibration. See paragraph 9.3.2
3.
Pump. See paragraph 9.3.3
4.
Alarm. See paragraph 9.3.4
5.
System. See paragraph 9.3.5
6.
Display. See paragraph 9.3.6
7.
Units. See paragraph 9.3.7
8.
Factory reset. See paragraph 9.3.8
Summary of Contents for LEO
Page 1: ...LEO Handheld Analyzer for IVF Applications Manual IST1446_REV05 SV 3 1 64 0 ...
Page 2: ...This page was left blank ...
Page 16: ...12 Figure 13 Pump ...
Page 84: ...80 Figure 110 T1 Calibration Factory Reset a b b a ...
Page 88: ...84 f e c a b d 2 1 1 Figure 114 CO2 O2 MODULE insertion ...
Page 92: ...88 f e c a b d 2 1 1 Figure 118 CO2 O2 MODULE insertion after pump replacement ...