24
4.
Put the cap on LEO gas inlet port with icon
(see Figure 28) and press on
Continue
(see Figure 29
a). Wait
stabilization
time (see Figure 29 b), during this time the pump is not running. Wait until the
Leakage evaluation
step is completed (see Figure 29 c).
3
Figure 28. Put the cap on LEO gas inlet port.
a
b
c
Figure 29. How to set Pump Leak Test
–
stabilization and Leakage evaluation steps (a
–
b
–
c).
If the Leak test is passed, the pump works properly. The green label showing
OK
appears as in Figure 30.