45
Note
►
If no device is selected, the data are stored on the current device shown on the homepage (1 in Figure
56 a).
Tip
►
During a gas measurement, you can start a temperature measurement and select a different Device
for which temperature data shall be stored.
2.
Wait for the temperature reading is stable.
3.
Press the
Temperature Logging
icon
(see Figure 56 a). The
Temperature Logging
settings page is
shown.
4.
Insert the measurement time using
+ and
–
(see Figure 56 b):
•
Each:
the time between two measurement points. It is the acquisition frequency. The time can be
inserted in minutes and seconds. Each time can be set at minimum of 5 seconds.
•
For:
the time interval of the Temperature Logging. The time can be inserted in hours and minutes.
5.
Enable/Disable the alarm on T1 and/or T2 (1 in Figure 56 c) and then press
Start
to start the
measurement (2 in in Figure 56 c).
Note
►
If T1 and/or T2 check boxes are enabled (see Figure 56 c), the alarm triggers when the reading
temperature values are out of the set range. See paragraph 9.3.4 to set the temperature alarm range.
Figure 56. Temperature logging (a
–
b
–
c)
When the Temperature Logging is in progress, the
Temperature Logging
icon becomes
(see Figure 57
b).
Note
►
During a Temperature Logging, the timer (1 in Figure 57 a) indicates the remaining time for
completing the time interval, set as “For” value (see
Note
►
The Temperature Logging can be interrupted before the time set by clicking on the icon
(see
a
b
c
1
1
2
Summary of Contents for LEO
Page 1: ...LEO Handheld Analyzer for IVF Applications Manual IST1446_REV05 SV 3 1 64 0 ...
Page 2: ...This page was left blank ...
Page 16: ...12 Figure 13 Pump ...
Page 84: ...80 Figure 110 T1 Calibration Factory Reset a b b a ...
Page 88: ...84 f e c a b d 2 1 1 Figure 114 CO2 O2 MODULE insertion ...
Page 92: ...88 f e c a b d 2 1 1 Figure 118 CO2 O2 MODULE insertion after pump replacement ...