11
Trap (1 in Figure 7). The end of Tube-B
(2 in Figure 11) must be connected to the sampling Gas Port of the
incubator.
2
1
Figure 11. Tube B
Note
►
Remove the protection caps from Tube-B before using it.
5.9
Additional tubes
Okolab supplies the following additional tubes (M in Figure 3):
1.
Tygon tube (OD = 6mm) with a length of 100 cm
2.
Tygon tube (OD = 4 mm) with a length of 30 cm
3.
Silicone tube (OD = 4mm) with a length of 10 cm
5.10
CO2-O2-MODULE (embedded in LEO)
The CO2-O2-MODULE is embedded into LEO, and it is included standard. It measures, with a high accuracy,
the CO2-O2 volume concentration in the unit of %, and CO2-O2 partial pressure in the units of mbar, psi and KPa.
CO2-O2-MODULE is temperature and pressure compensated thanks to a dedicated temperature and pressure
sensor that are included in this module.
The CO2-O2-MODULE is factory calibrated by using traceable standards, and stores the calibration data
in its internal memory. The declared accuracy is guaranteed for a period of 6 months from the factory calibration
date. The CO2-O2-MODULE is designed for an easy extraction from LEO (see paragraph 14).
The CO2-O2-MODULE lifetime is 5 years.
Figure 12. CO2-O2-MODULE
5.11
Pump (embedded in LEO)
The pump is a micro diaphragm pump that guarantee the desired gas flow rate (from 60 ml/min to 260
ml/min). The default flow rate is 150 ml/min. The pump is designed for easy extraction from LEO, see paragraph
18). Figure 13 shows an image of the pump inside LEO.
Summary of Contents for LEO
Page 1: ...LEO Handheld Analyzer for IVF Applications Manual IST1446_REV05 SV 3 1 64 0 ...
Page 2: ...This page was left blank ...
Page 16: ...12 Figure 13 Pump ...
Page 84: ...80 Figure 110 T1 Calibration Factory Reset a b b a ...
Page 88: ...84 f e c a b d 2 1 1 Figure 114 CO2 O2 MODULE insertion ...
Page 92: ...88 f e c a b d 2 1 1 Figure 118 CO2 O2 MODULE insertion after pump replacement ...