5-3
6600-0255-000
12/20/00
5/Maintenance
5-3
2.
Use a Phillips head screwdriver to remove the lens mounting screw located in the
center of the alarm light.
3.
Remove the bulb by pushing in and turning it counterclockwise.
4.
Install the new bulb by pushing in and turning it clockwise.
Note: Replace both bulbs for maximum lamp life.
5.
Place the lens cover in position and secure it with the mounting screw.
6.
Plug the power cord in and check for proper operation.
Figure 5-2
Alarm Lamp replacement
B. Observation Lamp Replacement
(See Figures 5-8, 5-9 and 5-10.)
Lamp: GE EXZ (Q50 MR16/NFL), Ohmeda Medical Stock No. 0208-0516-300 or GE
EXN (Q50 MR16/FL).
1.
Disconnect the power cord for the Warmer and allow the unit to cool for 10 minutes.
2.
Rotate the heater assembly to the side position (shown in Figure 5-3).
CI.05.077
Mounting Screw
Lens
Lamp
Push the alarm lamp
in and turn
counterclockwise to
remove it.