3-6
6600-0255-000
12/20/00
3/Operation
3-6
Check Patient Alarm:
1.
Manual Mode: The check patient alarm activates in the manual mode of operation
when the heater has been energized at a power level greater than 25% for 12
minutes. The alternating single tone alarm is activated for this condition. Pressing
the alarm silence switch silences this alarm and resets the timer for another 12
minutes of operation. If this alarm is not silenced within 3 minutes of the check
patient alarm occurring, an alternating two tone alarm is sounded and the heater is
deactivated until the alarm silence switch is pressed.
2.
Servo Mode: The check patient alarm activates in the servo mode of operation if the
heater has been at the 100% power level for 12 continuous minutes. The alternating
single tone alarm is activated for this alarm. Pressing the alarm silence switch
silences this alarm and resets the timer for another 12 minutes of operation. If this
alarm is not silenced within 3 minutes of the check patient alarm occurring, the
alternating two tone alarm is sounded and the heater is deactivated until the alarm
silence switch is pressed.
Power Failure Alarm:
A battery operated power failure alarm activates when the external power source fails or
is accidentally disconnected. The battery also provides power to an electronic memory to
recall previous control temperature settings for approximately 10 minutes, when the
battery is fully charged.
Table 1 — Warmer Alarms
Alarm Condition
Alarm Sound
Alarm Silence Period
Heater
Probe Failure‡
Alternating Two-Tone †
1 minute
Off
Patient Temperature
Alternating Two-Tone †
1 minute
Off
is greater than 42°C‡
Patient Temperature
Alternating Single-Tone †
5 minutes
*
more than 2°C from
Control temperature‡
Patient Temperature
Alternating Single Tone †
15 minutes
*
between 1 and 2°C §of
Control temperature‡
Patient Temperature
Alternating Two-Tone †
1 minute
Off
is less than 30°C ‡
System Failure**
Alternating Two-Tone
Cannot be
Off
silenced
Heat Off
Alternating Single Tone †
5 minutes
Off
Check Patient
Alternating Single Tone †
12 minutes
*
after 12 minutes
Alternating Two-Tone
15 minutes
Off
after 15 minutes
Power Failure
Alternating Single Tone
Cannot be silenced
Off
* Heater output is dependent on the Patient Temperature and Control Temperature
settings in the servo mode, and the % Power setting in the manual mode.
** Error code in elapsed time display indicates the cause of the alarm.
† Preceded by a 30 second operator prompt tone.
‡ Not applicable to model 2001 International
§ Can be set to between .5 and 2°C by qualified service personnel.
Summary of Contents for Ohio 2000 IWS Series
Page 1: ...Ohio Infant Warmer Systems All Models Operation and Maintenance Manual...
Page 6: ...iv 6600 0255 000 12 20 00 Table of Contents iv Notes...
Page 10: ...viii 6600 0255 000 12 20 00 Table of Contents viii Notes...
Page 12: ...x 6600 0255 000 12 20 00 Table of Contents x Notes...
Page 16: ...1 4 6600 0255 000 12 20 00 1 Description 1 4 Notes...
Page 24: ...2 8 6600 0255 000 12 20 00 2 Setup and Checkout Procedures 2 8 Notes...
Page 76: ...A 18 6600 0255 000 12 20 00 Appendix A 18 Notes...
Page 77: ...A 19 6600 0255 000 12 20 00 Appendix A 19 Notes...
Page 78: ...A 20 6600 0255 000 12 20 00 Appendix A 20 Notes...