A-3
6600-0255-000
12/20/00
Appendix
A-3
For your convenience, on Model 3000, 3100 and 3500 the proper bed-to-heater spacing is
indicated by a label located on the right aluminum upright (as viewed from the front).
Line Voltage Compensation:
Input line voltage is monitored and the heater drive is adjusted to compensate for
variations in the line voltage. This ensures a stable heater output despite input line
voltage fluctuations.
Circuit Breaker:
Rated Current: 7.5 A.
Trip Point: 9.45 A Minimum.
Type: Manual Resetting.
Model: Airpax Snapak.
Isolation Voltage:
2500 Vrms 60 Hz from the patient probe to the ac phase and neutral lines for one minute.
WARNING:
w
The patient probe is not isolated from earth ground. Any additional equip-
ment used with the Ohmeda Infant Warmer System must comply with UL
544, CSA 22.2, IEC 601, and VDE 750.
Leakage Current:
With ground wire open or connected and measured at an exposed metal surface, less
than 100 microamperes on 100 V and 120 V units (200 microamperes on 220 V, 230 V
and 240 V units).
Ground Resistance:
Less than 0.2 ohms as measured from the power cord plug
IEC 601-1 Specifications
Type of protection against electric shock:
Class 1
Degree of protection against electric shock:
Type B
m
Mode of Operation:
Continuous
Protection against hazards of explosion:
Not Protected
Protection against ingress of liquids:
Not Protected
Degree of electrical connection between
Equipment with an applied
equipment and patient:
part designed for non-conductive
connection to the patient.
Degree of mobility:
Fixed (Model 3050, 3100 and 3150)
Mobile (all others)
Controller
Electronics:
Microprocessor-based control system.
Self-test functions are performed at power up and during normal operation.
Power Control Method:
Proportional heat control with zero-voltage switching to minimize radiated and conducted EMI.
Observation Light:
Nominal illuminance output: 100 foot candles at center of mattress.
Estimated lamp life: 3,000 hours.
Summary of Contents for Ohio 2000 IWS Series
Page 1: ...Ohio Infant Warmer Systems All Models Operation and Maintenance Manual...
Page 6: ...iv 6600 0255 000 12 20 00 Table of Contents iv Notes...
Page 10: ...viii 6600 0255 000 12 20 00 Table of Contents viii Notes...
Page 12: ...x 6600 0255 000 12 20 00 Table of Contents x Notes...
Page 16: ...1 4 6600 0255 000 12 20 00 1 Description 1 4 Notes...
Page 24: ...2 8 6600 0255 000 12 20 00 2 Setup and Checkout Procedures 2 8 Notes...
Page 76: ...A 18 6600 0255 000 12 20 00 Appendix A 18 Notes...
Page 77: ...A 19 6600 0255 000 12 20 00 Appendix A 19 Notes...
Page 78: ...A 20 6600 0255 000 12 20 00 Appendix A 20 Notes...