CHAPTER 2 - DIAGNOSTIC GUIDE
Valor 1000 Series Service Manual
10
Ohaus Corporation www.ohaus.com
TABLE 2-2
DIAGNOSTIC GUIDE
Sympto
m
Possible Cause
Possible Solution
CAL E
Calibration error: unstable
environment or incorrect calibration
weight
Check environment and
calibration weights, and
then re-calibrate.
Err 8.1
The Power On weight on the
platform exceeds the zero range.
Clear pan.
Err 8.2
The Power On weight on the
platform is below the zero range.
Check if pan is missing.
Err 8.3
Over load (weight exceeds rated
capacity)
Remove weight from pan.
Err 8.4
Under load (pan removed)
Re-install pan.
Err 8.5
Tare weight out of range
Tare value exceeds
maximum.
Err 9.5
Internal data error or calibration data
error
Contact OHAUS or your
authorized dealer for help.
Err 13
Failure to write to EEPROM
Contact OHAUS or your
authorized dealer for help.
Lo.Bat
Battery very low
Replace batteries; Charge
rechargeable battery.
--NO--
Action not allowed
/
------
Busy
/