APPENDIX A
– SERVICE MENU
Valor 1000 Series Service Manual
30
Ohaus Corporation www.ohaus.com
APPENDIX A
– SERVICE MENU
This appendix describes Service Menu, and entry by following step:
1. Turn the balance off
2. Press and Hold the
On/Off
button and the
Tare
Button for 10 seconds. As the balance powers
up, appears followed by
Ramp
3. Select the function by pressing
On/Off
button, and change the function by pressing
Tare
Button.
4. Exit service menu with
End
A.1 Service Menu Structure
Function Display
Remark
Ramp
ramMp
Show Ramp value
Lin
LIN
Perform the Service linearity calibration
Span
SpaN
Perform the Service Span calibration
GEO
GEO
Hide/Unhide certain regions
Cap
Cap
Capacity
Grad
Grad
Gradacity
S.Unit
S.UNIt
Setup Unit
Filter
FIlter
Setup Filter
Expand
E.PAND
Expand reading
Lock
LOCK
Hide Calibration in the user menu
Lock the unit setting in user menu
Reset
reset
Reset
End
End
Exit servcie menu