CHAPTER 4 TESTING
Valor 1000 Series Service Manual
22
Ohaus Corporation www.ohaus.com
9.
Add the ten Delta x Delta values and divide by 9 Calculate the Standard Deviation by
applying the square root of the result from step 8.
Standard Deviation =___________
Note:
If the balance does not meet specifications, move it to a suitable location, ensure
that it is level, and try again.
4.2.3 Linearity Test
This test is used to determine the linearity of the unit throughout its operating range.
The masses used to perform this test can be utility masses
NOTE:
The scale must pass the Precision and Repeatability Tests, and be calibrated before
the Linearity Test may be performed.
TABLE 4-3. LINEARITY TEST MASSES
Capacity (g)
3000 x .5g
6000 x 1g
15000 x 2g
30000 x 5g
Reference Wt.
1000
1500
4000
8000
Load 1
500
1500
3000
6000
Load 2
1500
3000
9000
15000
Load 3
2000
4500
11000
22000
Capacity (g)
2000 x .2g
5000 x .5g
10000 x 5g
20000 x 10g
Reference Wt.
500
1200
2500
5000
Load 1
500
1200
2500
5000
Load 2
1000
2500
5000
10000
Load 3
1500
3800
7500
15000
NOTE:
All masses are nominal values. Use the same reference mass throughout the procedure.
1. Place the test mass on the Scale, record the weight and remove.
2. Place Load 1 on the Scale and press
Tare
3. Place the test mass on the Scale, record the weight and remove.
4. Place Load 2 on the Scale and press
Tare
5. Place the test mass on the Scale, record the weight and remove.
6. Place Load 3 on the Scale and press
Tare