THE BODY COMP SCALE 

BCS Series 

Operations Manual 

Summary of Contents for BCS Serie

Page 1: ...1 2 7 K i r k h a m R d P o w a y C A 9 2 0 6 4 P h 8 0 0 5 4 8 9 8 0 6 8 5 8 4 5 7 5 5 7 6 F a x 8 5 8 4 5 7 0 1 2 7 www bodycompscale com v a l h a l l a v a l h a l l a s c i e n t i f i c c o m TH...

Page 2: ...inter 5 2 5 Printouts 6 2 5 1 Plain Print Mode 6 2 5 2 Graphic Print Mode 6 SECTION III CODES LIMITS 3 1 Enabling Body Builder Ohms and Unit Count Functions 7 3 2 Weight Only Measurement 8 3 3 Limitat...

Page 3: ...ged externally internal damage may have occurred that may not be evident until operated If the unit fails to operate notify the carrier agent and Valhalla Scientific to obtain a return authorization n...

Page 4: ...nce we know the conductivity of someone s muscle mass we can accurately determine total body fat fat free mass and total body water 2 2 CONNECTIONS AND POWER UP Plug the AC adaptor 120VAC 10ACV or 220...

Page 5: ...hand makes contact with both handles on both sides of the display module The thumb should rest gently on top of upper probe The Body Comp Scale will induce a low frequency current for six seconds and...

Page 6: ...esting Energy Expenditure Daily Caloric Needs Model BCS 3 also will print Segmental Body Fat Displacement results 2 5 2 GRAPHIC PRINT MODE The graphic print mode is specifically design for the uses of...

Page 7: ...hen press the enter button Next In order to enable the Unit Count number to display on test reports This is available if you want to know exactly how many tests were administered 1 Remain on platform...

Page 8: ...dy fat percentage The typical ohmic value range for a healthy population Men 450 650 ohms Woman 570 850 ohms If you encounter a situation where the individual s ohms are not within the range mentioned...

Page 9: ...a sure weight on the platform A sure weight is one that has been previously determined The heavier the sure weight the more accurate the calibration 3 Adjust coarse screw CLOCKWISE to increase weight...

Page 10: ...scale and re test your body fat 5 Repeat steps 1 and 2 if additional adjustments are needed 4 3 CARE AND MAINTENANCE Proper care and cleaning of the Body Comp Scale is essential to ensure a long life...

Page 11: ...rized repair or damaged during shipment Valhalla Scientific must be notified promptly within the warranty period of the existence of a defect Valhalla Scientific shall have the sole right to determine...

Page 12: ...n body builders Possible Causes Super Capacitor is discharged and needs charging a Power adapter not plugged into the scale b Power adapter not plugged into outlet Scale was moved or picked up when pl...

Page 13: ...For additional information Call 1 800 548 9806 Visit our Website at www bodycompscale com 13...