7000 Series Indicators EN-25
5 D
3 D
2 D
1 D
StAbLE
0.5 D
rEAd
STABLE RANGE
HIGH
MEDIUM
FILtEr
LOW
rEAd
FILTER
3 D
1 D
0.5 D
AZt
OFF
rEAd
AZT
AUTO
ON
LIGHt
OFF
rEAd
BACKLIGHT
5 MINUTES
2 MINUTES
bLt.tmM
1 MINUTE
3.5.3
Stable Range
Set the amount the reading can vary while the stability symbol remains on.
0.5 d
= 0.5 graduations
1 d
= 1 graduation
2 d
= 2 graduations
3 d
= 3 graduations
5 d
= 5 graduations
NOTE:
When LEGAL FOR TRADE is set to ON, the setting is forced to 1 d. The setting is locked when
the Security Switch is set to the ON position.
3.5.4
Filter
Set the amount of signal filtering.
LOW
= faster stabilization time with less stability.
MEDIUM
= normal stabilization time with normal stability.
HIGH
= slower stabilization time with more stability.
3.5.5
Auto Zero Tracking
Set the automatic zero tracking functionality.
OFF
=
disabled.
0.5 D
= the display maintains zero until a drift of 0.5 graduation per second is
exceeded.
1 D
= the display maintains zero until a drift of 1 graduation per second is
exceeded.
3 D
= the display maintains zero until a drift of 3 graduations per second is
exceeded.
NOTE:
When Legal for Trade is set to ON, the setting is forced to 0.5 D. The 1 D and 3 D settings
are still available for applications that permit these settings. The setting is locked at the current
setting when the security switch is set to the ON position.
3.5.6
Backlight
Set the display backlight functionality.
OFF
= the backlight is always off.
ON
= the backlight is always on.
AUTO
= the backlight turns on when a button is pressed, or the displayed
weight changes and it turns off after the specified time period.
When AUTO is selected, set the time period.
1 MINUTE
= the backlight turns off after 1 minute.
2 MINUTES
= the backlight turns off after 2 minutes.
5 MINUTES
= the backlight turns off after 5 minutes.