EN-22 7000 Series Indicators
SEtUP
GRADUATION2
100
SEtUP
POWER UNIT
SEtUP
ZERO RANGE
SEtUP
AUTO TARE
A.tArE
ACCEPT
100
CUSTOM
TONNE
POUND-OUNCE
OUNCE
POUND
GRAM
KILOGRAM
ACCEPT
ON
5
2
1
0.5
3.4.8
Graduation2
Set the readability of Scale 2.
Settings of 0.0001, 0.0002, 0.0005, 0.001, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2,
0.5, 1, 2, 5, 10, 20, 50, and 100 are available.
NOTE:
Graduation2 setting selections are dependent on the Capacity2 setting. Selections are limited
to values that will provide a scale resolution between 1:1000 and 1:50000.
NOTE:
If DUAL was selected as the Range1 setting, the Graduation2 setting defines the fine range
graduation. The coarse range graduation is automatically defined as one step greater than the
Graduation2 setting.
3.4.9
Power Unit
Set the unit of measure displayed at startup.
AUTO
= last unit in use when the indicator was turned off.
KILOGRAM =
kilograms
GRAM
=
grams
POUND
= pounds
OUNCE
= ounces
POUND OUNCE = pound ounces
TONNE
= tonnes
CUSTOM
= custom unit
3.4.10
Zero Range
Set the percentage of scale capacity that may be zeroed.
2%
= zero up to 2 percent of capacity.
100%
= zero up to 100 percent of capacity.
3.4.11
Auto Tare
Set the automatic tare functionality.
OFF
= automatic tare is disabled.
ON
= the first stable gross weight is tared.
ACCEPT
= in Check weighing mode, stable gross loads within the accept limits
are tared.
When ACCEPT is selected, set the delay time.
OFF
= automatic tare takes affect as soon as the weight is stable.
0.5
= automatic tare takes affect after the weight is stable for 0.5 second.
1
= automatic tare takes affect after the weight is stable for 1 second.
2
= automatic tare takes affect after the weight is stable for 2 seconds.
5
= automatic tare takes affect after the weight is stable for 5 seconds.
Grad2
kg
0.0001
PwWr.UN
AUTO
ZErO
%
2
A.tArE
OFF
dELAY
OFF