7000 Series Indicators EN-29
UNIt
t
TONNE
ON
UNIt
t
OFF
UNIt
c
CUSTOM
ON
UNIt
c
OFF
UNIt
c
EXPONENT
-1
-2
3
2
1
UNIt
c
FACTOR
9.99999
UNIt
c
LSD
1000
UNIt
END UNIT
3.7.7
Tonne Unit
Set the status.
OFF
=
disabled
ON
=
enabled
NOTE:
Tonne Unit is not available when Range is set to DUAL or when Graduation setting is less
than 0.01 kilograms or 0.02 pounds.
3.7.8
Custom Unit
Use Custom Unit to display weight in an alternative unit of measure. The custom unit is defined using a conversion factor, where
the conversion factor is the number of custom units per kilogram expressed in scientific notation (Factor x 10^Exponent).
For example: To display weight in troy ounces (32.15075 troy ounces per kilogram) enter a Factor
of 3.21508 and an Exponent of 1.
Set the status.
OFF
=
disabled
ON
=
enabled
NOTE:
Custom Unit is not available when Range is set to DUAL.
When Custom Unit is set to ON, the Factor, Exponent and Least Significant Digit must be set.
Factor
Set the conversion factor using the numeric keypad.
Settings of 0.00001 to 9.99999 are available. The default setting is
1.00000
.
Exponent
Set the factor multiplier.
0
= multiply the Factor by 1 (1x10
0
)
1
= multiply the Factor by 10 (1x10
1
)
2
= multiply the Factor by 100 (1x10
2
)
3
= multiply the Factor by 1000 (1x10
3
)
-2
= divide the Factor by 100 (1x10
-2
)
-1
= divide the Factor by 10 (1x10
-1
)
Least Significant Digit
Set the graduation.
Settings of 0.00001, 0.00002, 0.00005, 0.0001, 0.0002, 0.0005,
0.001
, 0.002,
0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500 and 1000
are available.
NOTE:
Least Significant Digit setting selections are dependent on the Factor and Exponent settings.
Selections are limited to values that will provide a scale resolution between 1:1000 and 1:50000.
3.7.9
End Unit
Advance to the next menu or return to the top of the current menu.
e
0
FACtOr
0.00001
LSd
0.00001
Summary of Contents for T71P