Ranger™ Count 2000 Series EN-15
Power on unit
[
pwWr.UN
]
Set the unit of measure displayed at startup.
auto = last unit in use when turned off
kg
= kilograms
g
= grams
lb
= pounds
oz
= ounces
lb:oz = pound ounces
Zero Range
[
ZerO
]
Set the percentage of scale capacity that may be zeroed.
2%
= zero up to 2 percent of capacity
10% = zero up to 10 percent of capacity
Auto Opt
[
A.OPt
]
Set the automatic optimization functionality.
OFF = disabled
ON
= APW automatically optimized.
Auto Tare
[
A.tArE
]
Set the automatic tare functionality.
OFF
= Automatic Tare is disabled
ON
= the first stable gross weight is tared.
ON-ACC = stable gross loads within the accept limits are tared (in
Checkweigh mode)
Beeper Opt
[
bP.Opt
]
Set whether or not the beeper sounds when APW has been optimized.
OFF
= disabled
ON
= enabled
Beeper Signal
[
bp.sIG
]
Set how the beeper responds in Check Weighing mode.
OFF
= the beeper is disabled
ACCEPT = the beeper will sound when the weight is within the Accept range.
UNDER = the beeper will sound when the weight is below the Under setting.
OVER
= the beeper will sound when the weight is above the Over setting.
UNDER-
OVER
= the beeper will sound when the weight is below the Under setting
or above the Over setting.
Key Beeper
[
bp.KEY
]
Set whether the beeper sounds when a button is pressed.
OFF = no sound
ON
= sound