Summary of button navigation functions in menu mode:
--
Yes
Allows entry into the displayed menu.
- Accepts the displayed setting and advances to the next item.
--
No
Skips by the displayed menu.
- Rejects the displayed setting or menu item and advances to
the next available item.
--
Back
Moves backwards through the upper and middle level menus.
- Backs out of a list of selectable items to the previous middle
level menu.
--
Exit
Exits from menu directly to the active weighing mode.
4.2 Cal Menu
Enter this menu to perform calibrations.
Initial Calibration
When the scale is operated for the first time, a span calibration is recommended to
ensure accurate weighing results. Before performing the calibration, be sure to have
the appropriate calibration weights as listed in table 4-1. Ensure that the LFT
switch/calibration lock is set to unlocked position. See figure 5-1.
Or adjust the GEO setting according to your location (see table 4-2).
Procedure:
Press and hold
Menu
until [
mMeNU
] (Menu) is displayed. When the button is
released, the display will show [
C.A.L
]. Press
Yes
to accept. [
SpaN
] will then be
shown. Press
Yes
to begin the span calibration. [
0
kg] will be displayed. Press
Yes
to accept. [
--C--
] will be displayed while zero reading is stored. Next, the
display shows the calibration weight value. Place the specified calibration mass on
the pan. Press
Yes
to accept the weight or
No
to select an alternate weight. [
--C--
]
will be displayed while the reading is stored. The display will show [
done
] if the
calibration was successful. The scale returns to the previous application mode and
is ready for use.
TABLE 4-1.
Suggested Span Calibration Mass (sold separately)
Max
Mass
1
Max
Mass
1
1500g
1.5kg / 3lb
15000g
15kg / 30lb
3000g
3kg / 6lb
30000g
30kg / 60lb
6000g
6kg / 15lb
Note:
1
Pound masses are used when calibrating in the lb unit.