SERVICE MENU
Attention
: In case you have to replace the motor or the rotary encoder board
for any reason, a new calibration is required
urgently
.
Attention
: After all points have been reviewed and/or set up, you have to
press shortly the key start to store all parameters. Otherwise the setups were
all for nothing. If the storage has been successful in the display „rpm/rcf“ the
value of the imbalance sensor and the word „StorE“ are indicated.
2.1.13 Submenu Indication of the imbalance value of the rotor
Here you can read off the imbalance value of the rotor transponder.
• Activate the operation menu as described under „Activation of the operation
menu (Part 1)”.
• Turn the potentiometer, until the special digit „Y“ appears in the service
display.
• In the bottom line the real imbalance value is indicated. In the top line the
rotor number is indicated.
Attention
: To get access to this menu and read off the imbalance value, you
have to close the lid of the unit before „
Activate the operation menu“
.
2.1.14 Submenu Keyboard test
The keyboard test is used for checking the correct function of the foil keyboard.
• Activate the operation menu as described under „Activation of the operation
menu (Part 1)”.
•
Turn the potentiometer, until the special digit „P“ appears in the service
display.