GETTING STARTED
Page 11 of 53
SERVICEMANUAL
© OHAUS Corporation
Room temperature: 23°C
The absolute end temperatures can be subject to fluctuations of
±
2°C due to power tolerances
of the refrigeration unit.
The lowest sample temperatures are depending on the ambient temperature. If the ambient
temperature is increasing, the lowest sample temperature to be reached is increasing as well.
The acceleration times and deceleration times can be subject of small fluctuations, it depends
from unit to unit, and are therefore approximate values only.
1.7.4 Specifications
Model
FC5816
Speed Range
200 rpm -15000 rpm;10 rpm/set
Maximum RCF
21379 x g;10 x g/set
Maximum Capacity(Rotor)
6 x 250ml
Temperature range(N/A)
Air cool
Running Time
10 sec to 99 hr 59 min 59 sec or continuous
Noise level (depending on the rotor)
≤
61 ± 2 dB(A)
Allowable density at maximum speed
1.2 g/ml
Allowable kinetic energy
34363 Nm
Mains power connection AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
2.4 A
4.2 A
Power consumption
530 W
520 W
Dimensions (W × D × H)
446 x 538 x 354 mm
17.6 x 21.2 x 13.9 in
Net Weight (without rotor)
52 kg
115 lb
Shipping Dimensions (W × D × H)
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
Shipping Weight (without rotor)
77 kg
170 lb
Ambient conditions (EN/IEC 61010-1)
Environment
for indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
2°C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
Overvoltage category (IEC 60364-4-443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.
EMC
EN/IEC 61326-1 Class B emissions, Basic immunity
FCC Class B emissions