SERVICE MENU
Page 21 of 53
SERVICEMANUAL
© OHAUS Corporation
Attention
: In case that the signal generator is switch off, the keypad tone and
the audible error message will also be suppressed.
The keypad tone has to be switched on again after the reactivation of the
signal generator.
However the audible error message is automatically activated if the signal
generator has been reactivated.
Attention
: After all points have been reviewed and/or set up, you have to
press shortly the key start to store all parameters. Otherwise the setups were
all for nothing. If the storage has been successful in the display „rpm/rcf“ the
value of the imbalance sensor and the word „StorE“ are indicated.
2.1.9 Submenu Setup of the keypad tone
Here you can switch on resp. switch off the keypad tone.
• Activate the operation menu as described under „Activation of the operation
menu (Part 1)”.
• Turn the potentiometer, until the special digit „b“ appears in the service
display.
• By pressing the key rpm/rcf the display „rpm/rcf“ is now activated and you
can setup the respective option with the potentiometer.
• Now the keypad tone is either activated or not activated.
Attention
: In case that the signal generator is switched off, the keypad tone
and the audible error message will also be suppressed.
The keypad tone has to be switched on again after the reactivation of the
signal generator.
Attention
: After all points have been reviewed and/or set up, you have to
press shortly the key start to store all parameters. Otherwise the setups were
all for nothing. If the storage has been successful in the display „rpm/rcf“ the
value of the imbalance sensor and the word „StorE“ are indicated.