SERVICE MENU
Here you can adjust the different centrifuge types that are stored on the
control board.
• Activate the operation menu as described under „Activation of the operation
menu (Part 1)”.
• Activate the service menu as described under „Activation of the service
menu (Part 2)”.
• Turn the potentiometer, until the special digit „t“ appears in the service
display.
• Examples:
• By pressing the key rpm/rcf the display „rpm/rcf“ is now activated and you
can setup the respective option either with the potentiometer or by pressing
the key up/down
.
Attention
: Please do not adjust any other centrifuge types than the mentioned
one on the type label of the unit. Otherwise there may occur defects both in
the electronic and the mechanic components, too.
Attention
: After all points have been reviewed and/or set up, you have to
press shortly the key start to store all parameters. Otherwise the setups were
all for nothing. If the storage has been successful in the display „rpm/rcf“ the
value of the imbalance sensor and the word „StorE“ are indicated.
2.2.2 Submenu Adjustment of the operation mode
Here you can adjust the different operation modes stored on the control board.
• Activate the operation menu as described under „Activation of the operation
menu (Part 1)”.
• Activate the service menu as described under „Activation of the service
menu (Part 2)”.
•
Turn the potentiometer, until the special digit „C“ appears in the service
display. Do not confuse with „c“
Submenu Setup of the temperature.