MOUNTING SUPPORT
• In case the unit does not shut off you have to loosen the fixing screws of the
imbalance sensor. Adjust the imbalance sensor more sensitive with the
adjusting screw (clockwise).
• Repeat those steps as often as the listed value of the imbalance shut-off
table is reached.
• Switch off the unit with error message 01 and fill the rotor with the permitted
imbalance weight.
• The unit must not shut off after the start. In case the unit shuts off, you have
to lower the sensibility of the sensor.
• In case you have to readjust the imbalance sensor when testing the
permitted imbalance weight, you have to repeat the procedure with the
imbalance shut off.
• Turn the unit off and on -> the movement sensor is turned on again.
3.9 Replacing the control board
• Remove the lower covering of the front housing as described in chapter 3.1.
• Remove all electrical connections on the control board.
• Remove the 4 screws and remove the control board out of the unit.
• Put in the new control board, plug in all electrical connections and re-assemble the
unit in reverse order.
ATTENTION:
In case of replacing the control board you have to adjust the unit.
remove