EN-36
Defender 6000 Indicators
Project ID
[
P.ID
]
Set the Project identification number.
To set the number, short press the
No
button several times until the desired number appears. Short press
the
Yes
button to accept the number and move to the next digit. Repeat the process until all the digits are
correct. Press the
Yes
button to accept the value.
For i-DT61XWE model, you can input the Project ID through the numeric keyboard.
Scale ID
[
S.ID
]
Set the scale identification.
Please refer to the
Porject ID
section about how to set the number.
For i-DT61XWE model, you can input the Scale ID through the numeric keyboard.
End
[
End
]
Advance to the next menu or return to the top of the current menu.
4.7 Communication
Enter this menu to define external communication methods and to set printing parameters.
Data may be output to either a printer or PC.
Factory default settings are shown in bold.
4.7.1 RS232 Menu
Enter this menu to define communication parameters.
Note
:
2
nd
RS232 is only available for i-DT61XWE model.
Menu
Sub-Menu
Sub-Menu
(in segment) Options
Options (in segment)
RS232
(
r.S.2.3.2
)/
2nd SERIAL
PORT
(
2.S.E.r.I.A
)
Baud Rate
BAUd
300, 600, 1200, 2400, 4800,
9600
, 19200,38400, 57600
/
Parity
PArItY
7 Even, 7 Odd, 7 None,
8
None
7 EuEN
,
7 Odd
,
7 NONe
,
8 NONE
Stop Bit
StOP
1 bit
, 2 bit
/
Handshake
H.SHAKE
None
, Xon/Xoff
NONe
,
ONOFF
Alt Print CMD
ALt.P
'A' ~ 'Z',
P
/
Alt Tare CMD
ALt.t
'A' ~ 'Z',
T
/
Alt Zero CMD
ALt.Z
'A' ~ 'Z',
Z
/
Reset
reset
\
/
End
End
\
/
Reset
[
reset
]
Reset the RS232 menu to factory defaults.
NO
= do not reset.
YES
= reset
Baud Rate
[
BAUd
]
Set the baud rate (bits per second).
300
= 300 bps
600
= 600 bps
1200
= 1200 bps
2400
= 2400 bps
4800
= 4800 bps
9600
= 9600 bps
19200
= 19200 bps
38400
= 38400 bps
57600
= 57600 bps
Summary of Contents for Defender i-DT61PW