EN-30
Defender 6000 Indicators
Reset
[
reset
]
Reset the Setup menu to factory defaults.
NO
= do not reset
yes
= reset
Capacity Unit [
C.
UNIt
]
Select the unit used for calibration.
t (Metric Tonne)
g
kg
lb
Note:
t and g are not available for i-DT61XWE model.
Range
[
rANGE
]
Note:
This setting is only avaliable for i-DT61XWE model.
Set the number of weighing intervals.
This terminals can be configured to use single or dual interval. Each interval can be assigned its own
graduation. If dual interval is selected, the graduation will change when the weight reaches the second
interval.
When
Single
interval is selected, the additional parameters available are:
>|1|< Capacity
>|1|< Graduation
When
Dual
interval is selected, the terminal functions with two intervals, each with its own capacity and
graduation. In addition to the Interval 1 capacity and graduation parameters, the following two parameters
are available:
>|2|< Capacity
>|2|< Graduation
Capacity
[
Cap
] /
Capacity1
[
Cap1
]
Set the capacity of the scale or the capacity of the first scale (for i-DT61XWE model).
1…999999
Grad
[
GrAd
] /
Grad1
[
GrAd1
]
Set the scale readability or the readability of the first scale for i-DT61XWE model from 0.0001 to 100.
0.0001~100
Capacity2
[
Cap2
]
Set the capacity of the second scale for i-DT61XWE model.
Grad2
[
GrAd2
]
Set the readability of the second scale for i-DT61XWE model.
Power On Zero
[
P.ZerO
]
Zero the scale at Power On.
Off
= disabled.
ON
= enabled.
Power On Unit
[
P.UNIt
]
Set the unit that will be displayed at Power On.
AUtO
= last unit in use when turned off
kg
= kilograms
g
= grams
lb
= pounds
oz
= ounces
lb:oz
= pound ounces
t
= metric tonne
(
only available for i-DT61PW model
)